X-Rite eXact scanning spectrophotometer with full CC Capture support (Print, Color, Substrate and Instrument Inspectors). The quick report is shown on the instrument screen.
Hardware is not sold in the online store. Add products to your inquiry list and we will send you a quote by e-mail.
X-Rite Xact Advanced when working with CC Capture offers a lot of uniques features. The quick report is displayed on the instrument screen.
Perfect support for the following CC Inspectors:
Please note that X-Rite eXact can be configured, there are multiple models.
Important components that may change price are:
Please contact us for pricing details and we can assist you with options.
Please note: This offering is limited to the US market only.
eXact can be ordered with different Aperture 1.5, 2, 4, 6mm. The most popular Aperture is 2 mm and this is a typical choice for Offset printing, however, the user has to understand that aperture is related to screen frequency and minimal patch size.

For Print Inspector it is critical to capture data in a faster way - not by pointing each patch manually but by scanning it during movement. Optional trolley and components inside the instrument are required to use this feature. The instrument has to be prepared in the factory. Read X-Rite article about Scan option

ChromaChecker can offer eXact at a special price for bundles. Contact us and learn what you can get extra when ordering eXact.
| 3rd Party Software Supported | X-Rite ColorCert, X-Rite Color iQC, X-Rite InkFormulation, X-Rite NetProfiler and many others |
| Aperture | 1.5 mm, 2 mm, 4 mm or 6 mm |
| Battery | Lithium-Ion, 7.4VDC, 2200mAh |
| BestMatch Feature | Yes |
| Calibration | Automatic on integrated white reference |
| Color Differences | CIE ∆E* (1976), ∆ECMC, CIE ∆E* (2000), and CIE ∆E* (1994) |
| Color Spaces | CIE L*a*b*, CIE L*C*h° |
| Communication Interface | USB 2.0; Bluetooth (optional) - Class II |
| Configuration Options | Bluetooth, NetProfiler |
| Connectivity | USB 2.0; Bluetooth |
| Density Range | 0.0 D – 3.0 D |
| Density Standards | Status A, E, I, T and G |
| Dimensions (length, width, height) | 18 cm x 7.8 cm x 7.6 cm 7.1" x 3.1" x 3.0" |
| Embedded NetProfiler Support | Yes |
| Experience Level | Pressrooms, Press Operators and Production Supervisors looking to reduce make ready waste through tight color control. |
| G7 Certification | Yes with embedded G7, PSO, JapanColor job templates |
| Humidity | 30 to 85% RH (non-condensing) |
| Illuminants | A, C, D50, D55, D65, D75, F2, F7, F11, F12 |
| Import/Export Format | CxF3 |
| Industry Standards | ISO 5-4:2009(E), ISO 13655:2009 |
| Instrument Display | Color touch-screen, 320 x 240 pixels |
| Inter-Instrument Agreement | Average: 0.25 ΔEab, Max: 0.45 ΔEab (for M3: 0.55 ΔEab) (Measurements using X-Rite manufacturing standards at a temperature of 23°C +/- 1°C, 40-60% RH for all measurement modes on 12 BCRA color tiles and a white ceramic reference (D50, 2°)) |
| Internal Temperature & Humidity Sensor | 30 to 85% RH (non-condensing) |
| Languages Supported | Chinese Simplified, Chinese Traditional, Czech, Danish, Dutch, English, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Polish, Portuguese, Russian, Slovenian, Spanish, Swedish, Turkish, Vietnamese |
| Light Source | Gas-filled tungsten (illuminant type A) and UV LED |
| Macintosh | eXact Manager, DataCatcher and Driver: macOS 10.13, 10.14, 10.15, 10.16 (also on Apple M1 systems), eXact DataMeasure: MacOS X 10.13, 10.14 (32bit only) |
| Measurement Conditions | M0, M1 (part 1), M1 (part 2), M2, M3; also available in Xp variant for flexible packaging substrates |
| Measurement Geometry | 45°/0° ring illumination optics, ISO 5-4:2009(E) |
| Measurement Time | < 1 sec |
| NetProfiler Support | Yes |
| Observers | 2° and 10° |
| Operating Temperature Range | 10°C to 35°C 50°F to 95°F |
| PantoneLIVE Support | Yes; Embedded Pantone Libraries |
| Pass / Fail | Yes |
| Scanning Capability | Yes (Optional Upgrade) |
| Service Indicator | Certification Reminder |
| Short Term Repeatability - Density | '+/- 0.01 D for CMYK (Status E or Status T measurements, maximum error compared to mean value of 10 measurements every 5 seconds at 2.0 D except for M3 Yellow at 1.7 D) |
| Short Term Repeatability - White | 0.05 ΔEab (standard deviation) White BCRA (Error compared to mean value of 10 measurements every 5 seconds) |
| Software Development Kit | Yes |
| Software Modes | Basic |
| Spectral Analyzer | DRS spectral engine |
| Spectral Interval | 10 nm |
| Spectral Range | 400 nm to 700 nm |
| Spectral Reporting | 400 nm to 700 nm with 10 nm interval |
| Standards | More than 20'000 standards (incl. data for all M modes); New Color Library Tool to Create and Store Standards |
| Storage Temperature Range | -20° to 50°C -4° to 122°F |
| Supported Devices | Works with ColorCert, InkFormulation, NetProfiler |
| User Experience | Start-up wizard, configurable, customized toolset, upgradable |
| Voltage | 12 VDC @ 2.5 A |
| Warranty | 1 year |
| Weight | 700 g / 1.6 lbs. |
Collect everything you need in one inquiry list — we will get back to you by e-mail.