Please note that Myiro-1 is also supported by CC Display software and the user can use this instrument with Display Inspector and Lighting Inspector. We strongly advise using Myiro-1 as Master Instrument for Instrument correlation purposes.
![]() |
Quick Start (fast, easy and free) |
|
ChromaChecker offers a series of user-friendly procedures. This guides each user to easily perform basic tasks. Some Quick Start procedures can be performed without an active ChromaChecker account. This capability is provided for demonstration purposes only. CC Capture is G7® Certified. To use this tool in production requires a subscription. |
|
|
|
| Model | MYIRO-1 |
| Illumination/viewing system | 45°/0° (annular illumination)*1 |
| Spectral separation device | Concave grating |
| Wavelength range | Spectral reflectance: 380 to 730 nm Spectral irradiance: 360 to 730 nm |
| Wavelength pitch | 10 nm (5 nm - irradiance) |
| Half bandwidth | appr. 10 nm |
| Measurement area | ø 3.5 mm |
| Light source | LED |
| Measurement range | Reflectance: 0 to 150% / Density: 0.0 D to 2.5D |
| Repeatability | Colorimetric: Within ∆E00 0.05 (When the white calibration plate is measured 30 times in 10-second intervals after white calibration has been performed) |
| Inter-instrument agreement | Within ∆E00 0.3 (Average of 12 BCRA Series II color tiles compared to values measured with a master body under manufacturer's standard conditions) |
| Measurement time (single-point) | Appr. 1s |
| Measurement conditions *2 | M0, M1, M2, User-defined illuminant |
| Observers | 2° Standard Observer, 10° Standard Observer |
| Status indicator | LED to indicate instruments status |
| Interface | Wireless LAN (802.11 b/g/n); USB 2.0 |
| Scanning measurements | Scanning of color charts under all measurement conditions in one single scan |
| Power | USB bus power; Rechargeable internal battery |
| Dimensions (WxDxH) | 73 mm x 171 mm x 71 mm |
| Operating temperature/humidity range | 10° to 35°C, 30% to 85% relative humidity with no condensation |
| Storage temperature/humidity range | 0° to 45°C, 0% to 85% relative humidity with no condensation |
| *1 | Illumination for wavelengths under 400 nm is unidirectional |
| *2 | M0, M1, M2: Illumination conditions defined in ISO 13655 |
To enable a Wi-Fi connection, please install MYIRO-1 Configuration Tool Installer.
Wi-FI SSID and password must be uploaded to the instrument, and a wireless connection must be enabled.
Network tab:
Please note that your computer and Instrument must use the identical SSID.

A blue LED indicates a successful Wi-Fi connection.
To measure Lighting Conditions, an optional Ambient Light adapter is required. Before the first usage, it has to be registered using MYIRO-1 Configuration Tool Installer. The user needs to upload the adapter calibration data into the instrument memory.
Ambient Light Adapter tab:

"Current Serial No:" will report the related number when properly installed.
Additional information and Support Form is available for logged users.