Charge the Instrument before operation. Connect the supplied USB cable to the wall USB charger or computer. It can take a few hours for a full charge.

Please consult additional document: Understanding DS-700d
It is important to understand that the instrument requires two-step calibration
1. White Calibration - on the tile (found in the docking station)
2. Black Calibration- Instrument held in the vertical position at least 1m from ground. During Black calibration, random light should not enter the measurement port.
Connect the cable to the Instrument's USB-C socket. The base is dedicated to charging the internal battery - not connecting to the computer.
If You struggle with cable connection, make sure that a Spectrophotometer is listed on your USB device list. Do not replace the original cable with USB-C/USB-C - In such cases, the computer can't recognize the instrument! If you don't have a proper socket - use a converter or hub.

We have tested the USB connection, and all tested configuration works fine. Please let us know if you have experienced any issues. Please document details to help us solve the potential problems.
Activate Bluetooth in the instrument's menu. Please note that Instrument can be recognized by the dedicated software only, and it is not visible on the standard list of Bluetooth devices.
Next to the white calibration tile, there is a small label that reports factory-measured colorimetric references. Please note that Color Specification requires the following:
|
|
|
If an identical Color Specification is created in Color Inspector small difference can be observed, caused by small differences mainly in the D65 definition. There are two official = one based on the tabulated data and one based on a math equation (which enables any custom illuminant). The practical difference can be about 0.1 ∆E 00 and has to be accepted.
If you experience any issues with CC Capture, be sure to test a couple of things:

Additional information and Support Form is available for logged users.